|
Software Products for the Radiological Sciences
|
|
Required Beam Data
Use the generic machine and energies that we supply:
4, 6, 10, 15, 16, 18, and 24 MV x-ray beams. The difference in most
accelerators is due to the flattening filter.
But the effects of the flattening
filter are included in the beam field image that is measured,
as well as the in air collimator scatter factor which is also being
measured. The same applies for everything in the beam: wedges,
collimator jaws, MLC.
Dosimetry Check does not model the accelerator head. That is what is measured.
If the percent depth dose is close for two machines of the same energy,
phantom scatter is going to be close.
Persons using
generic data with various accelerators are getting good results. All you
have to do is define what a monitor unit is on your system at a single
point (specify the field size, SSD, depth, and dose rate in cGy/mu).
Just use it to test it.
Compute a 10x10 cm field on your planning system for a water phantom and take
a beam image of the 10x10 field.
Run DosimetryCheck to compare dose distributions. Repeat for a small and
large field size.
But if generic data is not satisfactory
enough, we can convert, or are willing to attempt to convert, your existing
beam data. The requirement is that you email us files that are in
ASCII format. We cannot read binary files. On an as need basis we
will write utilities to convert your data to our format. We may
all ready have the utility for your existing system. We need central
axis data, in air off center ratio on a diagonal, and in water off center
ratio on a diagonal for your largest field size, and output factors.
The program documentation gives more details.
Dosimetry Check uses a pencil beam algorithm.
Return to homepage
Math Resolutions, LLC
5975 Gales Lane, Columbia, MD 21045
support@MathResolutions.com
© copyright 2006 by Math Resolutions, LLC