Dosimetry Check Quick Start with Film


Print this page and use it to get you started. The manuals are online on the website for viewing and printing. Help screens are provided on all the toolbars and popups, just hit the Help button (see below if the Help screen cannot find the EnglishHelp.d directory).

  1. Film or a device? We have written a readers for the Varian aS500, Sun Nuclear Map Check, and PTW 729 ion chamber array. Contact us (see bottom of this page) if you want to use some other device. Any device used must be able to integrate the dose.

  2. We recommend you use 1 to 3 mm (3 for 18x) of copper to shield film from contamination electrons. Otherwise solid water will round off the fluence field edges a small amount as scatter within the solid water will produce the effect of a low pass filter.

  3. You will have to run a calibration curve covering the range from 0 mu to the largest you expect to encounter (and that the film can handle). You need to do this only once, as the curve can be later rescaled to correct for film processing. Just shoot a series using a 10x10 cm field size for the range of monitor units. We recommend you read about the step wedge as another method of easily producing a curve for each case.

  4. For what field size, SSD, and depth do you define the dose in cG/mu? You must tell the program in file "CalibrationXX", where XX is the energy, for each energy of your accelerator. Go into directory bd.d, go under the accelerator directory GeneralAccel, do an ls -F and note the photon energy directories there. Under each of those energies edit the calibration file if the definition is different from 100 cm SSD, 1.0 cG/mu at dmax. If you change an energy, you must run program tools.dir/ComputeCalConstant afterwards for the change to be incorporated in the beam model. [Run this program from a command prompt window, change directory to c:\mathresolutions on a Windows system]. See the section in the Dosimery Check manual on beam data for a description of the data in the beam data directory.

  5. Shoot a film of each treatment field. Using solid water will cause the effect of a low pass filter on the measured fluence, making a small difference in the end result as edges of sharp dose gradients will be round off slightly. We have not seen this effect using copper plate. Shoot an extra calibration film for each patient case.

  6. Your films will have to be digitized. We can read TIF, PNG, and Dicom. Most film scanners write to TIF format. Each of the above formats can handle 16 bit gray scale. We can also read images from the Radiological Imaging Technology system (Colorado Springs, CO, you need both the .hdr and .i00 file [.iv4 and .ir4 version 4]. You select either one with the mouse, including the .d00 file). If you have a file format we cannot read, email us an example file please.

  7. Download the plan in RTOG or Dicom RT format from your planning system. Select to receive the CT images, outlined regions of interest, 3D dose matrix, and beam geometry. (Corvus presently does not write the beam geometry in RTOG format and so you will have to create each beam and relocate isocenter and type in the beam gantry, collimator, and couch angles.)

  8. Run program ReadDicomCheck to read in plans written out in Dicom RT format. You must be able to navigate to the directory where the Dicom RT files are. You will need the CT scans, beam geometry, 3D dose matrix, and outlined regions of interest.
  9. OR Run the program ReadRtogCheck. This is an ASCII program, invoke it with:
    ReadRtogCheck -d directory_name
    from a command prompt window where you name the directory where the RTOG files are. ReadRtogCheck will create an entry in the patient directory, a stacked image set from the CT scans, and a plan using the plan name passed in the RTOG files.

  10. Next run DosimetryCheck. Select the patient, and then select the stacked image set. This is done from the first two pulldowns on the top main toolbar. A stacked image set is a set of images from which a 3D model of the patient can be generated. If you did not download the patient case from the planning system you will have to create a new patient and read in the CT scans to form a new stacked image set.

  11. Be sure your help screens come up. The location of the help directory is specified in the file rl.dir/English.d/HelpDirectory.loc and is presently in the directory rl.dir/English.d/EnglishHelp.d.

  12. A patient may have more than one stacked image set, which may or may not be fused together. Each plan created for a patient is associated with a stacked image set, the primary stacked image set. The primary stacked image set provides the skin boundary and the CT number to density conversion curve. The CT number to density curve is not passed in the RTOG or Dicom RT files. Go under the Stacked Image Set pulldown to Options, there select the stacked image set if you have more than one, and hit the Density button to get to the density curve toolbar. There you can select an existing curve, or create a new one. If you want to compare plans without inhomogeneity correction, create a straight line here so that all CT numbers map to 1.0 and select that curve.

  13. Again under Stacked Image Set -> Options select Skin and select the outlined region of interest that is the skin boundary. If this was not downloaded in the RTOG or Dicom RT files, than you will have to create one. Do this by returning to the main tool bar and select Contouring. You first have to create an outlined region of interest under the Volume pulldown, give it the name skin or body outline or something that denotes that. On the extreme right option menu select the Auto Body Contouring tool to create a body outline. When you use this tool the stacked image set must be the one displayed on the screen, as the tool uses the images displayed on the current screen (this allows for the flexibility of using reformatted images instead). Take a carefull look at all the frames in the stacked image set. If extra things were outlined you can do several things. Locate the couch top is one option. Another is to change the threshold for finding the body outline. If these do not work completely, you will have to delete or edit contours, after you accept the outlines. Be sure to accept the outlines. Any changes will now require the delete contour tool or the edit contour tool. See the System 2100 manual.

  14. Under the main toolbar go under Applications and select Field Dose. Here is where you read in the field images and convert using the calibration curve. The result for each field is stored in a file, which is then read later by each beam. See the Field Dose section in the Dosimetry Check manuel.

  15. Create a screen using the Screen Control button on the lower right hand corner of the main screen, and select a frame by hitting the button that occupies an empty frame. Under the Field Image pulldown select the images for your calibration curve. Under the Calibration pulldown select the Signal versus Dose tool. You use the measure tool to get signal values for the calibration curve tool. You measure the value at the center of the 10x10 cm field (or what ever field size you use for calibration) for the calibration here.

  16. Select Relative MU and enter a description. You must also specify the distance to the film. Then hit the Fit Data button. On the tool that comes up select the range for the polynominal fit. If the data is in terms of density use the range of 0 to 3 or higher. If in terms of transmitted light, try starting the range with a negative value and push in the Fit CD button.

  17. When you have a suitable curve, dismiss the fit tool and save the curve to a file. The curve should start with a zero value. The program will not extrapolate a polynominal fit, so you must cover the range of expected values.

  18. Notice the other options under Calibration. You can use the calibration curve to calibrate a step strip, Create Step Strip, (using a step wedge) and use the step wedge, Use Step Strip, in the future to generate a curve. Or you can rescale a curve to force agreement with a calbration film made with the patient case. You store the rescaled curve and use it in place of the original. To rescale, type in the signal value found on the test film and type in the correct monitor units. If you located the test field so that the distance is known, the measured value will show up in the Relative MU text field, otherwise it shows up only in the Dose text field and you have to formally type in the value.

  19. Read in the beam field images. Under Field Image select Locate Field for each of the beam field images. The tool is used for the current frame (that is outlined in red, click the mouse on a frame to make it current). Notice also the button in the upper right hand corner of each image which makes that frame full screen. You can resize the main screen any time. Operate the wheel controls to center the beam's eye view coordinate system on the beam field image. A tutorial exist for this function. You must locate the IEC beam's eye view coordinate system on each beam image. Remember that the beam's eye view system rotates with the collimator.

  20. You specify in the file Geometry the nominal angle for the accelerator collimator angle, in the accelerator's beam data directory. Note also the other coordinates specified in that directory. Beam position and angles are stored in an internal IEC coordinate system and the entries in the Geometry file used to translate those coordinates to the machine's coordinate system that you encounter in the beam positioning tools. The ReadRtogCheck and ReadDicomCheck programs translates the coordinates of beams into the internal IEC coordinate system, locating isocenter relative to the CT scans.

  21. After you have located the IEC coordinate system in all the beam images, you must next convert to relative monitor units using the above calibration curve. Select Convert To Dose under the Dose pulldown. Select the calibration curve. Make an image frame current by clicking the mouse on the frame. Select an empty frame to put the converted image into. If you need more frames hit Screen Control and either add more frames to your existing screen or add a new screen. Hit the Convert to Dose button. Notice that before you do this you can correct a result by specifying the mu used to make the film and the mu the beam is suppose to get in case you purposefully used less monitor units to shoot the film. Leave the pixel size at 0.1 cm. After converting all the beam image files, you must save them. Make each result of the conversion current and select the Save Dose Field under the Dose pulldown. Give each beam file a meaningful name, such as beam1 or left_lat_50.

  22. You have done the work. The rest is easy. Return to the main toolbar and under Applications select Dosimetry Check. Select the Plan. Because this system is multi-plan, you have to specify where the plan is to be displayed. On the Display pull down you can select a default screen for a transverse, coronal, and sagittal image through isocenter or the center of an ROI. You can always go under the Display pulldown on the Plan toolbar to select to display in the current frame or screen. You can make reformatted images under the pulldown on the main tool bar under the Stacked Image Set pulldown. Or you use any of the images displayed for the stacked image set.

  23. Under the Plan tool bar select the first beam. Under Options on the beam toolbar, select Read Field. A screen will be created to hold the images of the beam files. Select the file where you stored the converted field image above and hit the Apply button. The program will display the image, then increment the beam tool bar to the next beam. [Note you can go directly to any beam's toolbar using the option menu at the left on the current beam toolbar.]

  24. If you did not get the beam positions with the RTOG or Dicom RT files, you will have to create each beam. The successive beams will start with the isocenter of the prior beam. You will have to type in the gantry, collimator, and couch angle under the Move pulldown on the beam toolbar.

  25. Be sure the right energy is selected for each beam. That should come up correctly with RTOG and Dicom RT.

  26. Return to the plan toolbar from the beam toolbar. Under Evaluate select Compare 2d Dose if you got a 3D dose matrix from the planning system. Make a frame current where the plan is being displayed on a 2d image. On the top of the compare tool select what you want to see among dose from Dosimetry Check, foreign dose (from the planning system), and dose difference. Note only one thing can be tinted however. Type in a dose value and hit the enter key. Hit the Display in Current Frame button at the bottom of the popup.

  27. If you did not get a 3D dose matrix, then under Evalute select Display Dose in Current Frame and then select 2D Isodose Lines. In this case use the same image planes used with the plan and reproduce the same isodose levels for comparison of hard copy.

  28. To calculate the dose to specific points, return to the main toolbar, go under the Stacked Image Set pulldown to Options, and then select Points. Locate the points with the mouse on images of the stacked images set. Return to Evaluate under the plan toolbar and select Point Doses. On the point dose toolbar, select display, print, or file points. Compare the dose to the point for what the planning system computed. Note here you can locate an isocenter point by specifying the coordinates in beam's eye view, which would be (0,0,0).

  29. To get a hard copy of any image, click the mouse on the image and then hit the Print Screen button on your key board.

Return to Dosimetry Check Page
Return to homepage